# Scanning Capacitance Microscopy (SCM)

> Maps charge carrier polarity/distribution in semiconductors.

Source: https://covalent.com/techniques/more-techniques/scanning-capacitance-microscopy-scm/
Updated: 2026-04-23T00:01:17+00:00

---

![](https://spcdn.shortpixel.ai/spio/ret_img,q_cdnize,to_auto,s_webp:avif/covalent.com/wp-content/themes/covalent/assets/images/blog-default-banner.jpg)

 ![](https://spcdn.shortpixel.ai/spio/ret_img,q_cdnize,to_auto,s_webp:avif/covalent.com/wp-content/themes/covalent/assets/images/blog-mobile-banner.jpg)

 [Core Lab](https://covalent.com/covalent-core-lab/)

# Scanning Capacitance Microscopy (SCM)

SCM is a scanning probe microscopy technique that maps charge carrier polarity and distribution in semiconductors with sub-100nm resolution. It evaluates carrier concentrations from 1e16-1e20 cm⁻³ through dC/dV signals, ideal for SiC-MOSFET and IGZO-TFT analysis.

 [Add to Quote](#)

## What Is Scanning Capacitance Microscopy (SCM)?

Scanning Capacitance Microscopy (SCM) is a specialized scanning microscopy technique that provides detailed information about charge carrier in semiconductors with sub-100nm resolution.

SCM evaluates carrier concentration by measuring the local capacitance, namely dC/dV, under an applied AC voltage between the conductive AFM tip and the sample. This technique simultaneously acquires AFM images, enabling direct correlation between electrical properties and device structure for comprehensive semiconductor analysis.

Resolution

### Resolution

Sub-100nm spatial resolution for carrier mapping in semiconductors.

Analytical Depth

### Analytical Depth

Maps carrier polarity and concentration beneath the device surface.

Correlated Imaging

### Correlated Imaging

Combines AFM topography with electrical data for complete device insight.

[https://covalent.com/wp-content/uploads/2026/01/measurement-of-the-local-capacitance.jpg](https://covalent.com/wp-content/uploads/2026/01/measurement-of-the-local-capacitance.jpg)

The above graph shows the measurement of the local capacitance under an AC bias voltage to the sample using an AFM probe; the carrier polarity and carrier concentration can be evaluated through the dC/dV signal.

## Why Use SCM?

- Advanced sample preparation techniques support high data reliability.
- Although STEM-EDX helps detect [impurities](/solutions/contamination-analysis/) with a 1e21 cm⁻³ concentration or more, SCM is necessary for concentrations below 1e20 cm⁻³.
- Evaluates charge carrier polarity and concentration with a sub-100nm resolution.

[https://covalent.com/wp-content/uploads/2026/01/Cross-sectional.webp](https://covalent.com/wp-content/uploads/2026/01/Cross-sectional.webp)

Cross-sectional SCM image showing carrier distribution in a semiconductor device. The dC/dV signal differentiates between n-type (blue, negative dC/dV) and p-type (pink/red, positive dC/dV) regions beneath the gate metal structure. Scale bar represents 1μm, demonstrating SCM’s ability to map carrier polarity and spatial distribution with high resolution.

### Analytical Depth

Carrier distribution maps and semi-quantified carrier concentration line profiles.

### Material Compatibility

Semiconductors with carrier concentrations of 1e16 cm⁻³ to 1e20 cm⁻³.

### Depth Profiling

Any depth region can be accessed using a cross-sectional sample.

## Covalent’s Capabilities Offer SCM for High‑Resolution Semiconductor Carrier Mapping

[Get a Quote](https://covalent.com/get-a-quote/)

## Working Principle

SCM utilizes an AFM probe to measure local capacitance changes under AC bias voltage. The technique generates dC/dV signals that directly correlate with carrier polarity and concentration. Cross-sectional specimens enable depth profiling in any region of interest. Simultaneous AFM imaging provides structural correlation with electrical properties. Advanced sample preparation ensures reliable data from 1E16 to 1E20 cm⁻³ carrier concentrations.

### Equipment Used for SCM:

#### Bruker Dimension XR (SCM option), NanoScope V

- AFM platform: Large-sample SPM with PeakForce Tapping for routine atomic-defect and sub-nanometer imaging in air and liquid.
- Nanomechanics: AFM-nDMA for quantitative nanoscale storage/loss modulus over 0.1 Hz–20 kHz, linked to bulk DMA.
- Nanoelectrical: DataCube modes for hyperspectral electrical mapping, giving electrical and mechanical spectra at every pixel.
- Nanoelectrochemistry: PeakForce SECM / EC-AFM for <100 nm-resolution electrochemical imaging with simultaneous topography, electrical, and mechanical contrast.

[View Spec Sheet](https://www.bruker.com/en/products-and-solutions/microscopes/materials-afm/dimension-xr-afm.html)

### Key Differentiators

SCM provides direct feedback on fabrication processes for impurity injection verification. The technique offers exceptional sensitivity for low-concentration carriers below 1e20 cm⁻³ where other methods fail.

#### Strengths

- Detects carrier polarity and detects small amounts of charge carriers.
- Provides direct feedback on the fabrication process.
- Highly reliable experimental data due to Covalent’s advanced technique.

#### Limitations

- Not for [metals](/applications/metals-testing-and-failure-analysis-lab/) or insulators.
- Best resolution: 20 nm (decreases at low concentration); max area 80×80 μm².
- Higher carrier sensitivity, but lower spatial resolution than STEM-EDX.

Two colleagues working together on a laptop, discussing a project in a bright, professional environment.

## Unsure Whether SCM Is Right for You?

SCM delivers sub-100nm carrier mapping for semiconductors with concentrations from 1e16-1e20 cm⁻³ and verifies that your fabrication process achieved the designed impurity injection.

[Talk to an Expert](https://covalent.com/contact-us/)

## Sample Information

Example Outputs

Sample Requirements

[https://covalent.com/wp-content/uploads/2026/01/SCM-SiC-MOSFET-output.jpg](https://covalent.com/wp-content/uploads/2026/01/SCM-SiC-MOSFET-output.jpg)

Carrier distribution in SiC-MOSFET and semi-quantified carrier concentration profile along the arrows shown in the figure.

[https://covalent.com/wp-content/uploads/2026/01/SCM-IGZO-TFT-output.jpg](https://covalent.com/wp-content/uploads/2026/01/SCM-IGZO-TFT-output.jpg)

The figure above shows the carrier distribution in IGZO-TFT for evaluating the relationship between gate length and channel length.

### What we accept:

- Semiconductor sample (patterned device or wafer).
- Semiconductors with carrier concentrations of 1e16 cm-3 to 1e20 cm-3.
- The sample area of interest should be larger than 0.5×0.5 μm2.
- Cross-sectional sample preparation is necessary for SCM measurements.

## Use Cases

- Semiconductor
- Power Electronics

- ![Semiconductor](https://spcdn.shortpixel.ai/spio/ret_img,q_cdnize,to_auto,s_webp:avif/covalent.com/wp-content/uploads/2025/12/Semiconductor.jpg)

### Semiconductor

SCM enables precise metrology of p, p+, n, and n+ regions by measuring their width and thickness with nanoscale accuracy. The technique evaluates charge carrier distribution patterns critical for device performance optimization. SCM proves invaluable in [failure analysis](/solutions/failure-analysis/) and reverse engineering by identifying dopant migration, junction degradation, and process variations.
- ![Power Electronics](https://spcdn.shortpixel.ai/spio/ret_img,q_cdnize,to_auto,s_webp:avif/covalent.com/wp-content/uploads/2025/12/Power-Electronics.jpg)

### Power Electronics

SCM characterizes wide-bandgap semiconductors like SiC and GaN in power devices that operate at high voltages and temperatures. The technique maps carrier concentration gradients in MOSFET channels and evaluates junction integrity in IGBT structures. SCM supports reliability testing by detecting early-stage degradation mechanisms in power semiconductor components.

## Complementary Techniques

- **SIMS:** For impurity dopant concentration evaluation along with the depth direction.
- **STEM-EDX:** Used to evaluate impurity spatial distribution.
- **SIMS and SCM:**SIMS is often combined with SCM to consider the activation rate of impurities.

## Why Choose Covalent for Your SCM Needs?

Covalent delivers SCM testing with unmatched sensitivity and spatial resolution for precise carrier polarity and distribution evaluation. Our expertise extends beyond standard measurements, including optimized experimental conditions and advanced cross-sectional specimen preparation techniques. This comprehensive approach ensures highly reliable data acquisition and superior analytical results for your semiconductor characterization needs.

## Frequently Asked Questions

Identifying the right test can be complex, but it doesn’t have to be complicated.
 Here are some questions we are frequently asked.

- ### Is it possible to quantify carrier concentrations in any type of semiconductor?

Semi-quantification is possible for Si and SiC semiconductors using our proprietary calculation models. Qualitative carrier distribution mapping and polarity identification remain available for other semiconductor materials.
- ### What measurement error can be expected with SCM carrier concentration analysis?

The error ranges from half an order of magnitude to one order of magnitude of carrier concentration.
- ### Can SCM differentiate between edge and screw dislocation types in materials?

No, SCM cannot.
- ### What minimum feature size can SCM resolve in patterned semiconductor devices?

Roughly 50-100 nm.
- ### How does ambient humidity affect SCM measurement accuracy and reliability?

SCM data is collected in the temperature/humidity-controlled laboratory room. We don’t think that the humidity is a critical problem.
- ### Can SCM analyze buried interfaces in multilayer semiconductor device structures?

SCM can detect if the carrier concentration in the region of interest is higher than in other areas.
- ### How accurately can SCM measure individual semiconductor layer thickness in heterostructures?

SCM can if the thickness of each layer is more than several tenth of nms

Resources

 ![](https://spcdn.shortpixel.ai/spio/ret_img,q_cdnize,to_auto,s_webp:avif/covalent.com/wp-content/themes/covalent/assets/images/resource-bg-c.svg)
Resources

 ![](https://spcdn.shortpixel.ai/spio/ret_img,q_cdnize,to_auto,s_webp:avif/covalent.com/wp-content/themes/covalent/assets/images/resource-bg-o.svg)

 [![Beyond the Surface: Next-Generation Scanning Acoustic Microscopy](https://spcdn.shortpixel.ai/spio/ret_img,q_cdnize,to_auto,s_webp:avif/covalent.com/wp-content/uploads/2026/03/next-generation-scanning-acoustic-microscopy-768x430.jpeg)

Webinar

Beyond the Surface: Next-Generation Scanning Acoustic Microscopy

57 mins

Apr 16, 2026](https://covalent.com/webinars/next-generation-sam/)

 [![PED and DPC](https://spcdn.shortpixel.ai/spio/ret_img,q_cdnize,to_auto,s_webp:avif/covalent.com/wp-content/uploads/2026/04/covalent-ped-and-dpc-case-study-768x973.png)

Case Study

PED and DPC

Apr 08, 2026](https://covalent.com/resource-library/ped-and-dpc/)

 [![Seeing the Unseen: The Power of Electron Microscopy](https://spcdn.shortpixel.ai/spio/ret_img,q_cdnize,to_auto,s_webp:avif/covalent.com/wp-content/uploads/2026/02/the-power-of-electron-microscopy.jpg)

eBook

Seeing the Unseen: The Power of Electron Microscopy

Feb 25, 2026](https://covalent.com/resource-library/the-power-of-electron-microscopy/)

 [![Photoinduced Force Microscopy (PiFM) for Photonics Analysis](https://spcdn.shortpixel.ai/spio/ret_img,q_cdnize,to_auto,s_webp:avif/covalent.com/wp-content/uploads/2026/01/photoinduced-force-microscopy-pifm-for-photonics-analysis.jpg)

Case Study

Photoinduced Force Microscopy (PiFM) for Photonics Analysis

Jan 19, 2026](https://covalent.com/resource-library/pifm-for-molecular-identification-of-contamination-on-photomasks/)

 [![Solving Bond Uniformity Challenges with SAM](https://spcdn.shortpixel.ai/spio/ret_img,q_cdnize,to_auto,s_webp:avif/covalent.com/wp-content/themes/covalent/assets/images/resource-thumbnail.jpg)

Video

Solving Bond Uniformity Challenges with SAM

Dec 30, 2025](https://covalent.com/resource-library/solving-bond-uniformity-challenges-with-sam/)

 [![Failure Analysis in Action: Fracture Analysis of Biomedical Appliances with Advanced Microscopy](https://spcdn.shortpixel.ai/spio/ret_img,q_cdnize,to_auto,s_webp:avif/covalent.com/wp-content/themes/covalent/assets/images/resource-thumbnail.jpg)

Video

Failure Analysis in Action: Fracture Analysis of Biomedical Appliances with Advanced Microscopy

Dec 17, 2025](https://covalent.com/resource-library/failure-analysis-in-action-fracture-analysis-of-biomedical-appliances-with-advanced-microscopy/)

Need Expert Guidance?

Our experts can help you determine if this approach is the best fit for your samples and discuss the next steps.

[Talk to a Materials Scientist](https://covalent.com/contact-us/)