SEM Detector Types: How to Choose the Right Detector

Learn about the different types of detectors on SEM tools as well as four operational ideas to remember to optimize your SEM analysis.

An SEM image is only as good as the signal behind it. Secondary electron (SE) detectors, such as the Everhart-Thornley detector (ETD) and the through-lens detector (TLD), show surface topography. Backscattered electron (BSE) detectors, such as the CBS, DBS and in-column detectors, show differences in composition. This guide explains the main SEM detector types and where they sit in the microscope. It then covers four things to check before you image: which detectors you have, where your contrast comes from, your accelerating voltage and your detector angle.

SE vs BSE: The Two Signals SEM Detectors Collect

SignalWhere it comes fromWhat it showsTypical detectors
Secondary electrons (SE)Low-energy electrons knocked out of the top few nanometers of the sampleSurface topography, edges, cracks and fine texture. Raised features and edges appear brighter.ETD with positive grid bias, TLD, ICE, T2, T3, LVD
Backscattered electrons (BSE)Beam electrons scattered back out of the sample, from deeper in the materialComposition: heavier elements appear brighter, lighter elements darker. Low-angle BSE also carries topography.CBS, DBS, ICD, MD, T1, ETD with negative grid bias

Where SEM Detectors Sit in the Microscope

The electron beam (1) travels down the column and focuses on the sample. In-column and through-lens detectors (2), such as the TLD and Trinity detectors, sit inside the column. They collect electrons that travel back up through the lens, which gives the highest resolution. Below-lens BSE detectors (3), such as the CBS and DBS, are ring-shaped and sit just under the pole piece, catching BSE that leave the sample at high angles. In-chamber detectors (4), such as the Everhart-Thornley detector, are mounted at an angle on the side of the chamber and pull in secondary electrons with a biased grid. The sample (5) emits both signals at once, so the detector you choose decides what the image shows.

Diagram of Covalent detector components showing electron beam, in-column detector, below-lens BSE, E.
Covalent detector schematic illustrating SE/EDS pathways and detection components in a SEM.

SEM Detector Types

In-chamber detectors

DetectorFull nameSignalBest for
ETDEverhart-Thornley detectorSE with positive grid bias; high-angle BSE with negative grid biasGeneral imaging. Found on most full-size SEMs.
ICEIn-Chamber Electron detector (Thermo Scientific)SE and secondary ionsSimilar to the ETD, with higher sensitivity and a wider grid bias range. Used for electron and ion imaging on DualBeam systems.
DBSDirectional Backscatter detector (Thermo Scientific)BSELow kV and low beam current imaging at fast scan speeds. Each of its segments can be combined or subtracted.
CBSConcentric Backscatter detector (Thermo Scientific)BSERetractable ring detector under the pole piece. Inner rings give composition contrast, outer rings add topography, and it works well at low kV.
LVDLow Vacuum DetectorSEImaging in low vacuum mode, for samples that charge or outgas

In-column and in-lens detectors

DetectorFull nameSignalBest for
TLDThrough-Lens DetectorSE or BSE, depending on suction tube biasUltra-high-resolution imaging in immersion mode, and deep holes with the suction tube voltage at maximum
ICDIn-Column DetectorBSE, plus SE with stage biasSolid-state detector for high-resolution composition contrast
MDMirror DetectorBSESingle-segment BSE in immersion or non-immersion mode, with beam deceleration and stage bias
T1Trinity 1BSEIn-lens BSE imaging, including at tilt, in high or low vacuum
T2Trinity 2SE [confirm]High-resolution surface imaging [confirm]
T3Trinity 3SEInsertable in-column detector with the highest resolution and lowest signal

At Covalent, the Helios 5 UC DualBeam uses the TLD, ICD and MD in-column detectors. The Scios DualBeam uses the Trinity system (T1, T2, T3) and the LVD.

4 Things to Remember When Choosing an SEM Detector

Figure 1: Both the Helios 5 UC DualBeam and Scios DualBeam contain multiple detectors to optimize critical signal gathering
Figure 1: Both the Helios 5 UC DualBeam and Scios DualBeam contain multiple detectors to optimize critical signal gathering

1. Which detectors does your SEM have?

SEM detectors can be mounted in the sample chamber, below the pole piece or inside the column, and the set varies from instrument to instrument. Each detector has its own ideal conditions (high or low vacuum, high or low kV, fast or slow scan speed) and collects a different mix of signals. Some specialize in secondary electrons, others in backscattered electrons. Knowing which detectors are available, and how your settings affect each one, helps you refine the measurement and build a complete picture of the sample. The detector tables above are a quick reference.

2. Where is your contrast coming from?

Contrast comes from how efficiently each signal scatters, and without enough of it, features and defects can be lost. SE and BSE create contrast in different ways, so decide first what you need to see. Secondary electrons are surface-sensitive and are produced more at bumps, ridges, cracks, facets and fine texture, so those areas appear brighter. With BSE, materials with higher atomic numbers scatter more electrons and appear brighter, while lighter materials appear darker.

Figure 2: The same sample imaged with two detectors, showing different contrast. From Covalent Academy webinar #18, “Expert Lessons on Electron and Ion Beam Imaging and Interpretation.”

3. Is your accelerating voltage penetrating below the surface?

Accelerating voltages typically range from 1 to 30 kV, depending on how sensitive the sample is, the vacuum mode and the information you need. Higher voltages penetrate deeper into the sample. They give better resolution, are less affected by aberrations and are easier for new users, but they can hide surface detail. Lower voltages bring out fine surface features and surface composition, but resolution drops and they are harder to work with, especially below 2 kV. As a rule, choose a high voltage for material contrast and a low voltage for surface detail. Many projects image at several voltages to get the full picture.

Figure 3: SE images of Tantalum hand on a barium/carbon/oxygen oval on a cracked iron/tantalum/silicon substrate taken with different accelerating voltages. From Covalent's Covalent AcaBSE images of the same area at different accelerating voltages. BSE gives high resolution with strong material contrast: dark means low atomic number and bright means high atomic number. From Covalent Academy webinar #18.demy webinar #18.

Higher resolution, with excellent materials contrast. Image can reliably be interpreted as: dark = low atomic number, bright = high atomic number.
Higher resolution, with excellent materials contrast. Image can reliably be interpreted as: dark = low atomic number, bright = high atomic number.

Figure 4: BSE images of Tantalum hand on a barium/carbon/oxygen oval on a cracked iron/tantalum/silicon substrate taken with different accelerating voltages. From Covalent's Covalent Academy webinar #18.

4. What angle is your detector at?

A detector’s position decides which signal it picks up. The angle at which electrons leave the sample depends on the beam energy, the sample itself and any electrical bias on the stage. BSE collected at high angles, close to the beam, depend less on the surface and carry more composition information. BSE collected at low angles, closer to the sample, are more sensitive to topography. SE detectors such as the ETD are mounted at an angle and use a biased grid to attract more secondary electrons.

Diagram of detector positions inside a DualBeam SEM chamber and column
Figure 5: Detector positions on the DualBeam systems at Covalent.

Which SEM Detector Should You Use?

You want to seeSignalDetector and settings
Surface topography and fine textureSETLD in immersion mode, or the ETD, at low to moderate kV
Composition and phase differencesBSECBS, DBS, ICD or T1, using high-angle segments
Beam-sensitive or charging samplesSE or BSELow kV with the DBS or CBS, or the LVD in low vacuum
Deep holes and trenchesSETLD with the suction tube voltage at maximum
Ion images on a DualBeamSE and ionsICE
Elemental compositionX-raysEDS (see SEM-EDS)

SEM Analysis with Covalent Live View

Covalent Live View lets you join your sample’s analysis session live with a Covalent expert, wherever you are. You see what our scientists see and can discuss the project as it unfolds. With Live View you can:

  • Understand the analysis in full context
  • Make sure the critical features are captured by imaging or EDS
  • Work with the measurement expert to adjust the analysis in real time
  • Invite colleagues to review measurements and next steps together

Live View is available with SEM analysis at no extra charge.

SEM Testing Services at Covalent

Covalent’s SEM team is led by Dr. Avery Green [confirm], who has worked extensively with scanning electron microscopy and focused ion beam (FIB) analysis. Our Helios 5 UC DualBeam gives the resolution needed for devices and complex nanostructures. Our Scios DualBeam offers comparable ultra-high resolution and adapts to difficult materials, including samples that were previously hard to image in an electron microscope. Our scientists choose the detectors, voltages and imaging conditions for each question and walk you through the results. For techniques we don’t run in-house, we can place work through Covalent Connect, our network of trusted partner labs.

Turnaround is typically 5–7 business days, with rush options available at an extra cost.

Frequently Asked Questions

What are the main types of SEM detectors?

Secondary electron (SE) detectors, such as the Everhart-Thornley detector and the through-lens detector, image surface topography. Backscattered electron (BSE) detectors, such as the CBS, DBS and in-column detectors, show composition differences. Most SEMs also carry an EDS detector for elemental analysis.

What is an Everhart-Thornley detector?

The Everhart-Thornley detector (ETD) is the standard SE detector on most SEMs. It sits at an angle in the chamber and uses a biased grid, a scintillator and a photomultiplier. With a positive grid bias it collects secondary electrons; with a negative bias it collects only high-angle backscattered electrons.

What is the difference between SE and BSE imaging?

SE imaging shows surface shape and texture, because secondary electrons come from the top few nanometers. BSE imaging shows composition, because heavier elements backscatter more electrons and appear brighter.

What is a CBS detector?

A Concentric Backscatter (CBS) detector is a retractable, ring-shaped BSE detector mounted under the pole piece. Its concentric segments can be used separately: inner rings for composition contrast and outer rings for topography.

What is a through-lens (TLD) or in-lens detector?

A through-lens or in-lens detector sits inside the column and collects electrons that travel back up through the final lens. It gives the highest resolution, especially in immersion mode, and can image deep holes and trenches.

How do I choose the right accelerating voltage?

Use higher voltages (up to 30 kV) for material contrast and resolution, and lower voltages (below 5 kV) for surface detail and beam-sensitive samples. Many analyses combine several voltages.