ICP-MS Analysis: Inductively Coupled Plasma Mass Spectrometry

ICP-MS analysis measures metals and trace elements down to ppt levels in liquids, chemicals, and digested solids, making it the most sensitive routine technique for elemental impurities.

What Is Inductively Coupled Plasma Mass Spectrometry (ICP-MS)?

Inductively coupled plasma mass spectrometry (ICP-MS) is an elemental and isotopic analysis method. A high-temperature argon plasma ionizes the sample. A mass spectrometer then separates and counts those ions, giving both concentration and isotopic information at ultra-trace levels.

Labs use ICP-MS to measure trace metals in aqueous solutions, process chemicals and digested solids. Detection limits reach the ppt range for most elements and approach ppq on our multi-quadrupole system. ICP-MS measures dozens of elements at once and keeps a linear response across many orders of magnitude, so one run can cover both major and trace constituents.

Wide Material Range

Works with liquids, digested solids and solid samples by laser ablation across industries.

Ultra-trace Detection

Measures elements at ppt levels, with ppq sensitivity for selected elements.

Isotopic Precision

Resolves isotope ratios for tracer studies, contamination sourcing and research.

Why Use ICP-MS?

ICP-MS is the right choice when a result depends on very low concentrations. It detects trace and ultra-trace elements that ICP-OES, XRF and EDS cannot see. That makes it the standard method for elemental impurity testing, trace metal contamination and isotope ratio measurement.

In semiconductor manufacturing, ICP-MS checks the purity of process chemicals and wafers. Pharmaceutical teams use it to show compliance with ICH Q3D and USP <232>/<233>. It also supports environmental monitoring, battery and energy materials, biomedical research and forensic work.

If your samples have higher concentrations or heavy matrices and you don’t need isotopes, ICP-OES may be faster and less expensive. For metal contamination on wafer surfaces, see VPD-ICP-MS.

Triple-Quadrupole ICP-MS

Removes interferences for accurate sub-ppt quantification.

Laser Ablation Capability

Analyzes solids directly with micrometer-scale mapping and depth profiling.

Flexible Workflows

Supports automation, chromatography coupling and high-throughput batches.

How ICP-MS Analysis Works

Sample preparation comes first. Solids are usually digested in acid, often by microwave digestion, and liquids are diluted to a suitable matrix level. A nebulizer (1) then sprays the prepared solution into a fine mist.

The mist enters an argon plasma (2) that runs at several thousand degrees. The plasma dries the droplets, breaks down the sample and ionizes each element, mostly to singly charged (+1) ions. The sampler and skimmer cones (3) pull those ions out of the plasma and into the vacuum system as a focused beam.

A collision/reaction cell (4) uses gases such as helium, oxygen or ammonia to remove polyatomic interferences that would otherwise overlap the element of interest. Finally, the quadrupole mass filter (5) sorts ions by mass-to-charge ratio (m/z) and a detector counts them. The result is a concentration for each element and, when needed, its isotope ratios.

With laser ablation, a laser replaces the nebulizer. It removes material from a solid surface and a carrier gas sweeps it into the plasma, so solids can be analyzed and mapped without digestion.

Diagram of how ICP-MS works: nebulizer, argon plasma, interface cones, collision/reaction cell, and mass filter with detector
How ICP-MS works: the sample is nebulized (1), ionized in an argon plasma (2), extracted through the interface cones (3), cleaned of interferences in the collision/reaction cell (4), and sorted by m/z and counted (5).

Equipment Used for ICP-MS​

At Covalent, we pair the Thermo Scientific iCAP TQ ICP-MS and the PerkinElmer NexION 5000 with the Teledyne Photon Machines IRIDIA laser ablation system. This setup covers routine liquid workflows, the most demanding ultra-trace work and direct analysis of solids.

Thermo Scientific iCAP TQ ICP‑MS

  • Triple‑quadrupole interference removal: TQ technology with a Collision/Reaction Cell to strip away complex matrix interferences, enabling confident sub‑ppt elemental quantification.
  • High sensitivity with routine usability: Provides the lowest detection limits in Thermo’s benchtop ICP‑MS line while still supporting simple method development through Qtegra™ ISDS software and Reaction Finder tools.
  • Consistency at scale: Engineered for uptime and repeatability, minimizing downtime and ensuring consistent, accurate results even over large sample batches.
  • Flexible coupling with advanced sampling devices: Compatible with automation, chromatography, and laser ablation, expanding analysis modes beyond solution nebulization.
  • Direct solid analysis with IRIDIA laser ablation: Fast washout (<1 ms), homogeneous energy delivery, and variable spot sizes allow high‑resolution elemental mapping and depth profiling of solid materials without full digestion.
Specifications

PerkinElmer NexION® 5000

  • Industry-first four-quadrupole design: Full-length Q0+Q1+Universal Cell+Q3 with <0.7 amu resolution for interference removal beyond triple-quad.
  • Sub-ppt detection limits: BECs <1 ppt even in hot plasma for ppq-level sensitivity in semiconductor, biomonitoring, and advanced materials.
  • Universal Cell with dynamic bandpass tuning: Reactive and collision gas modes for precise elimination of complex polyatomic interferences.
  • Maintenance-free plasma system: 34 MHz LumiCoil™ RF generator + OmniRing™ Triple Cone Interface for unmatched matrix tolerance and long-term stability.
Specifications

Key differentiators

Strengths

  • Exceptional sensitivity, with ppt detection limits for most elements.
  • Isotopic discrimination for isotope ratio and tracer studies.
  • Wide dynamic range, from ultra-trace up to percent-level concentrations in one method.
  • Broad material compatibility, from high-purity chemicals to digested ceramics, polymers and biological samples.
  • Spatial resolution and depth profiling of solids through laser ablation.
  • Flexible workflows with automation, chromatography coupling and laser ablation.

Limitations

  • Most samples need conversion into a solution (excluding laser ablation); prep can be difficult depending on element solubility and stability.
  • Accuracy requires calibration with matrix-matched standards.
  • Isotopic measurements can be affected by polyatomic interferences.
  • Limited to elemental analysis; provides no molecular or bonding information.
  • Surface specificity is limited; depth profiling only possible with laser ablation.
  • Technique depends on careful sample prep and interference management.

Example Outputs

Each ICP-MS run produces mass spectra and calibration curves. Together they show which elements are present and at what concentration.

Mass spectrum: This high-resolution scan covers 105 to 155 amu. Each peak is a different element or isotope. Even at ultra-low concentrations, ICP-MS gives clear peaks with very little background, which acts as a fingerprint of the sample’s elemental makeup.

Calibration curve: This curve for ¹⁷⁵Lu is linear across more than 10 orders of magnitude. ICP-MS stays accurate from trace amounts up to much higher concentrations, so major and minor elements can often be measured in the same run.

Covalent turns this raw data into clear reports. You receive:

  • A list of all detected elements, plus isotopes if they are part of the study.
  • Concentrations in ppt, ppb or ppm, corrected for digestion and dilution.
  • Interpretation from our scientists, tied to your application or regulatory question.

Mass Spectrum: Here’s a high-resolution scan across the 105–155 amu range. Each peak represents a different element or isotope in the sample. Even at ultra-low concentrations, ICP-MS can pull out a clear signal with very little background noise; this acts as the “fingerprint” of the elemental makeup.

Calibration Curve: The image below shows a quantification curve for ^175Lu. The straight-line relationship across more than 10 orders of magnitude demonstrates just how wide the dynamic range is. Simply put, ICP-MS can measure incredibly tiny amounts and still stay accurate all the way up to much higher concentrations

Sample Requirements

Good sample preparation decides whether ICP-MS data is accurate or inconclusive.

  • Acceptable forms: liquids, and solids after digestion or by laser ablation.
  • Sample amount: a few milligrams of solid or a few milliliters of liquid is often enough. 
  • Key properties: solubility, volatility and stability of the analytes guide the preparation route.
  • Special considerations: organics, ceramics and refractory materials need tailored digestion. High-salt samples are diluted to protect the plasma and reduce matrix effects.
  • Contamination control: for ultra-trace work, we recommend clean, acid-leached PFA or PP containers. 

 

Covalent reviews each sample individually and tailors the preparation so results stay accurate at the lowest levels.

ICP-MS vs ICP-OES: Which Do You Need?

ICP-MS and ICP-OES both use an argon plasma. ICP-MS measures the ions, while ICP-OES measures the light the atoms emit. Choose ICP-MS when you need ppt detection limits, isotope ratios or a full trace impurity screen. Choose ICP-OES when concentrations are in the ppb to percent range, the matrix is high in dissolved solids, or cost per sample matters more than detection limits.

 ICP-MSICP-OES
Typical detection limitsppt (ppq for some elements)ppb
Isotope ratiosYesNo
Matrix toleranceLower, about 0.2% dissolved solidsHigher, several percent dissolved solids
Best forUltra-trace impurities, high-purity materialsMajor and minor elements, routine QC

ICP-MS Applications by Industry

Mining & Geology

ICP-MS detects platinum group metals like Pt, Pd, Rh, and Ir at sub-ppb levels in rocks and ores. This allows geologists to accurately assess ore value and plan extraction strategies that would be missed with less sensitive methods.

Oil & Petrochemical Refining

Refineries use ICP-MS to monitor trace elements such as Ni, V, and rare earth metals in fuel oils and petrochemical streams. Even minimal concentrations can impact catalyst lifetimes, process efficiency, and regulatory compliance, making precise detection critical.

Pharmaceuticals

ICP-MS supports ICH Q3D and USP <232>/<233> compliance by measuring elemental impurities such as arsenic, cadmium, mercury and lead well below their permitted daily exposure limits. This protects patients and supports quality assurance throughout drug development.

Environmental Monitoring

ICP-MS detects toxic elements such as arsenic, lead and uranium in water and soil at very low levels. When coupled with chromatography (HPLC-ICP-MS), it can separate harmful inorganic arsenic from less toxic organic forms.

Semiconductor & Advanced Materials

Sub‑ppt detection of alkali and transition metals is essential for high-purity wafers and chemicals in chip fabrication. ICP-MS reveals trace contaminants that can reduce yields or damage device performance. Laser ablation capabilities further allow high-resolution mapping of solid samples and precise trace element analysis in challenging matrices.

Techniques That Complement ICP-MS

Measures carbon, hydrogen, nitrogen, sulfur and oxygen, which ICP-MS cannot quantify well.

Add to Quote
GD-MS

Ideal for ultra-trace bulk analysis in solids. Complements ICP-MS when extreme sensitivity for metals in difficult matrices is required.

Add to Quote

Uses the same plasma principle but measures optical emission instead of ions. ICP-OES is less sensitive than ICP-MS but tolerates heavier matrices and costs less per sample. ICP-MS adds isotopic analysis, which ICP-OES cannot do.

Add to Quote

Sensitive to halides and small organic acids, which are outside the typical measurement scope of ICP-MS.

Add to Quote

Offers lower sensitivity (~0.1 wt%) but can be non-destructive. SEM-EDS can measure elements such as carbon, oxygen, nitrogen, and halides that are difficult for ICP-MS, though light elements like Li and Be are more easily quantified with ICP-MS.

Add to Quote
Surface-sensitive and non-destructive, XPS measures elemental composition and provides molecular or bonding information. Sensitivity is lower (~0.1 at%), but it gives chemical context that ICP-MS cannot.
Add to Quote
XRF

Gives rapid, non-destructive elemental analysis of solids with minimal preparation. Sensitivity is lower than ICP-MS, and accuracy depends on sample homogeneity.

Add to Quote

Collects metal contamination from a wafer surface by vapor phase decomposition and measures it by ICP-MS. Use it when you need surface metals in atoms/cm² rather than bulk concentrations.

Add to Quote

ICP-MS Testing Services at Covalent

Covalent runs ICP-MS in our core lab on the Thermo Scientific iCAP TQ and the PerkinElmer NexION 5000, with IRIDIA laser ablation and microwave digestion in-house. That combination lets us adapt the method to more sample types and matrices than a standard setup, with cleaner spectra and lower detection limits.

Our scientists review every sample before analysis. They plan the digestion, matrix handling and interference removal that decide whether ICP-MS data holds up. You get reliable numbers and a clear explanation of what they mean for your material, process or regulatory filing.

Turnaround is typically 5–7 business days, with rush options available at an extra cost.

Frequently Asked Questions

What is ICP-MS used for?

ICP-MS measures trace and ultra-trace elements in liquids and solids. Common uses include elemental impurity testing of pharmaceuticals, purity checks of semiconductor chemicals and wafers, environmental monitoring, battery materials and isotope ratio studies.

What is the difference between ICP-MS and ICP-OES?

Both use an argon plasma. ICP-MS detects ions by mass and reaches ppt levels with isotope information. ICP-OES detects emitted light, reaches ppb levels and tolerates heavier matrices. ICP-MS is the choice for ultra-trace work, and ICP-OES for higher concentrations and routine QC.

What are the detection limits of ICP-MS?

For most elements, detection limits are in the low ppt range in clean solutions. Some elements reach ppq levels on multi-quadrupole systems. Real limits depend on the sample matrix, dilution and blank levels.

Can ICP-MS be used for isotopic ratio analysis?

Yes. ICP-MS separates isotopes of the same element by mass, so it can measure isotope ratios with high precision. This supports tracer studies, contamination source tracking and isotope dilution quantification.

What is the difference between quadrupole and ToF ICP-MS?

A quadrupole mass spectrometer filters ions so that only one mass‑to‑charge ratio (m/z) is measured at a time. In contrast, a ToF analyzer separates ions by their velocity after acceleration through an electric field, allowing detection of a wide range of m/z values simultaneously.

How do I prepare high-salt samples for ICP-MS?

Dilute high-salt samples well before analysis, usually to below about 0.2% total dissolved solids. This reduces matrix effects and protects the cones and plasma. Our team can recommend dilution or matrix removal for your sample.

How does laser ablation ICP-MS work for solid samples?

A focused laser beam ablates tiny particles from the surface of a solid sample. These particles are transported to the RF plasma by a carrier gas, at which point they are ionized, separated, and measured in the same way as a solution sample in ICP‑MS.

Is ICP-MS used for ICH Q3D elemental impurity testing?

Yes. ICP-MS is the most common technique for ICH Q3D and USP <232>/<233> testing, because it measures all the listed elements well below their permitted daily exposure limits in one method.