Raman vs FTIR Spectroscopy: Differences and When to Use Each

FTIR and Raman spectroscopy both identify materials from their molecular vibrations, but they detect different bonds and work best on different samples. FTIR is the first choice for most unknown organics. Raman reaches smaller features, works through glass and water, and reads crystal phase and stress. Many projects use both. This page compares the two and explains how our scientists choose between them.

Raman vs FTIR at a Glance

FTIRRaman
What it measuresInfrared light absorbed by the sampleLaser light scattered by the sample at shifted energies
Strongest forPolar groups: C=O, O–H, N–HNon-polar bonds (C–C, C=C), carbon materials, crystal phases
Smallest feature at CovalentAbout 10 µm (micro-FTIR)About 1 µm (spatial resolution better than 0.5 µm)
Water and glassWater absorbs stronglyWorks through glass and in water
Main limitThin films under about 100 nm are hard to detectFluorescence can swamp the signal; pure metals give little or no signal
Typical usesPolymer ID, adhesives, organic residuesGraphene and carbon, polymorphs, stress mapping

How FTIR Works

FTIR shines infrared light on the sample and measures which wavelengths are absorbed. Each absorption band matches a molecular vibration, so the spectrum works as a fingerprint that can be searched against reference libraries. Our lab measures FTIR in transmission, reflection and ATR modes, and with an IR microscope for small features.

How Raman Spectroscopy Works

Raman focuses a visible or UV laser on the sample and measures the small fraction of light that scatters back at a shifted energy. Each shift matches a molecular or lattice vibration. Because a confocal microscope collects the signal, Raman can measure features down to about 1 µm and map how composition, phase or stress changes across a surface.

Key Differences Between Raman and FTIR

Which bonds each technique sees

A vibration shows up in FTIR when it changes the molecule’s dipole, which is why polar groups such as C=O and O–H give strong bands. A vibration shows up in Raman when it changes how easily the electron cloud deforms, which favors non-polar bonds such as C–C, C=C and S–S. In molecules and crystals with a center of symmetry, a vibration seen in one technique is often silent in the other. That’s why pure silicon is hard to measure by FTIR but gives a strong Raman peak.

Feature size

Micro-FTIR measures features from about 10 µm. Raman measures features down to about 1 µm, which makes it the better choice for single particles, thin lines and small defects.

Water, glass and sample preparation

Water absorbs infrared light strongly, so wet samples are hard to measure by FTIR. Water and glass give weak Raman signals, so Raman can measure through a glass vial or in solution. Both techniques measure most samples as received, in air.

Limits

FTIR has trouble with films thinner than about 100 nm and with conductive materials. Raman struggles with fluorescent samples, which can hide the signal, although a different laser wavelength often helps. Pure metals give little or no Raman signal, but oxides and coatings on metals can be measured.

When to Choose Each Technique

Choose FTIR first when

  • The sample is an unknown organic material, such as a polymer, adhesive, oil or residue
  • You need a library match to name the material
  • The feature or contaminant is larger than about 10 µm
  • You need IR transmission of an optical film, filter or window

Choose Raman when

  • The feature is smaller than about 10 µm
  • The sample is wet or sealed in glass
  • The question is about crystal phase, polymorphs, stress or strain
  • The material is carbon-based, such as graphene, carbon nanotubes or diamond-like carbon
  • You need to map composition or phase across a surface or a wafer up to 300 mm

When to Use Both

Mixtures and layered parts often need both techniques. A filled polymer, for example, may show the resin clearly in FTIR and the inorganic filler in Raman. Running both on the same spot removes guesswork about what each band means.

The whitepaper shows this with polystyrene measured by both techniques. Some bands appear only in FTIR, others only in Raman, and the bands they share have different relative intensities. Read together, the two spectra give a fuller picture of the structure than either one alone.

[Image: Figure 4c, polystyrene measured by FTIR and Raman, with bands unique to each technique highlighted]

Equipment Used for FTIR and Raman

ThermoFisher Scientific Nicolet iS50 FT-IR Spectrometer

  • Spectral range: 7800 to 350 cm⁻¹
  • Resolution: better than 0.09 cm⁻¹
  • Wavenumber accuracy: better than 0.005 cm⁻¹
  • Multi-detector configuration, including a temperature-controlled DLaTGS detector for stable intensity measurements
  • Diamond ATR (standard and high-pressure) for hard samples, and Ge ATR for carbon-filled or opaque materials
  • Variable-angle ATR and transmission for thin film analysis
  • Reflectance mode for coatings and contaminants on metal and other reflective surfaces
  • Beam-defining apertures that reduce stray IR radiation for accurate transmission (FTIR-optical) measurements

ThermoFisher Scientific Nicolet Continuum IR Microscope

  • 15× and 32× objectives
  • Sampling modes: transmission, reflection, Ge ATR and diamond-cell transmission
  • Dedicated ATR objective with interchangeable diamond, Ge, ZnSe and Si crystals and an integrated pressure sensor
  • Grazing-angle objective (NA 0.99) for thin films and surface layers
  • Dual detector bay for MCT-A, MCT-B and InGaAs detectors

Oxford Instruments WITec360 Semiconductor Edition

  • Confocal Raman and photoluminescence (PL) microscope designed for advanced materials and semiconductor characterization
  • Excitation sources: 355 nm, 532 nm
  • 300 mm wafer capability: motorized vacuum wafer stage supports automated mapping and characterization of semiconductor wafers up to 300 mm in diameter
  • TrueSurface™ topography tracking: maintains optimal focus across rough, curved, patterned and non-planar samples by automatically following surface topography
  • Vibration-isolated platform: enables high stability measurements
  • Configurable spectral resolution: multiple spectrometer grating options can support both wide spectral range acquisitions and high-resolution characterization

ThermoFisher Scientific DXR3xi Raman Spectrometer

  • Multiple excitation lasers: 455 nm, 532 nm, 785 nm
  • Laser power with precision controls: 0.1 mW power increments
  • Spatial resolution: better than 0.5 micron
  • Confocal depth resolution: better than 2 micron
  • Maximum image area: 101.6 mm x 76.2 mm

Sample Requirements

FTIR

  • Transmission: films ideally under 10 µm thick
  • Transmission and reflection: flat surfaces give the best data quality
  • Micro-FTIR: features from about 10 µm to 150 µm, on samples up to about 20 mm tall
  • All forms: solids and liquids are accepted, as long as they are stable under infrared light

Raman

  • Solids, powders, films, fibers, particles and liquids, mostly as received, in air with no vacuum or coating
  • Feature size: down to about 1 µm. Wafers: up to 300 mm
  • Samples must tolerate laser exposure. Sensitive or dark materials are run at reduced power
  • Highly fluorescent samples may need a different laser wavelength (355, 455, 532 or 785 nm)

What’s in the Whitepaper

  • How FTIR and Raman each characterize organic and inorganic systems
  • Why some materials, such as pure silicon, are hard to measure by FTIR, and where Raman fills the gap
  • A side-by-side FTIR and Raman comparison of polystyrene, with band assignments
  • Uses in device and packaging R&D: identifying unknowns, comparing research samples, mapping how uniform the composition is, and surface chemistry

FTIR and Raman Testing Services at Covalent

Our scientists choose FTIR, Raman or both based on your sample and the question you need answered, then interpret every spectrum against our reference libraries. When one technique isn’t enough, we pair them with GC-MS, SEM-EDS, XPS or XRD in the same lab to explain what changed and why. Each report includes annotated spectra, the identified materials and what the results mean for your process, product or failure.

Turnaround is typically 5–7 business days, with rush options available at an extra cost.

Frequently Asked Questions

Is Raman better than FTIR?

Neither is better overall. FTIR is usually the first choice for unknown organics because polar groups give strong, library-searchable spectra. Raman is better for features under about 10 µm, for wet samples or samples in glass, and for crystal phase, stress and carbon materials.

What is the difference between an FTIR spectrum and a Raman spectrum?

An FTIR spectrum shows how much infrared light the sample absorbs at each wavenumber. A Raman spectrum shows how much laser light scatters back at each energy shift. Both are plotted in cm⁻¹ and both reflect molecular vibrations, but strong bands in one are often weak or missing in the other.

Can FTIR and Raman be run on the same sample?

Yes. Both measure most samples as received and in air, so we can run them on the same spot. That’s often the fastest way to identify mixtures, filled polymers and layered parts.

Why is water a problem for FTIR but not for Raman?

Water absorbs infrared light strongly, so its bands cover the spectrum of whatever is dissolved or wet. Water gives only a weak Raman signal, so Raman can measure samples in solution or through a glass container.

Which technique can analyze smaller features?

Raman. It measures features down to about 1 µm, compared with about 10 µm for micro-FTIR.

Can Raman analyze metals?

Pure metals give little or no Raman signal. Oxides, corrosion products and coatings on metals can be measured.

How long does FTIR and Raman analysis take?

Each measurement takes minutes. Project turnaround is typically 5–7 business days, with rush options available.