TGA vs DSC: Thermal Characterization Techniques Explained

TGA measures how a material’s mass changes as it heats. DSC measures the heat flowing in or out. Here’s what each reveals, and when to use both.

TGA vs DSC in One Line Each

Thermogravimetric analysis (TGA) weighs a sample as it heats and shows what leaves it: moisture, solvents, additives and decomposition products. Differential scanning calorimetry (DSC) measures heat flow into or out of a sample and shows what changes inside it: glass transition, melting, crystallization and cure.

TGA vs DSC at a Glance

TGADSC
MeasuresMass change vs temperatureHeat flow vs temperature
RevealsDecomposition, moisture, volatiles, filler or ash contentGlass transition (Tg), melting, crystallization, cure
Typical sample10–50 mg5–10 mg
Range at CovalentAmbient to 1200 °C−90 to 550 °C
Answers“How much of it is there, and when does it break down?”“When does it soften, melt or cure?”

When to Run Both

Run both when a polymer behaves unexpectedly. DSC shows whether the Tg or crystallinity has shifted, and TGA shows whether the filler, plasticizer or moisture content has changed. Running TGA first also sets a safe upper temperature for the DSC run, so the sample doesn’t decompose inside the DSC cell.