TGA vs DSC in One Line Each
Thermogravimetric analysis (TGA) weighs a sample as it heats and shows what leaves it: moisture, solvents, additives and decomposition products. Differential scanning calorimetry (DSC) measures heat flow into or out of a sample and shows what changes inside it: glass transition, melting, crystallization and cure.
TGA vs DSC at a Glance
| TGA | DSC | |
|---|---|---|
| Measures | Mass change vs temperature | Heat flow vs temperature |
| Reveals | Decomposition, moisture, volatiles, filler or ash content | Glass transition (Tg), melting, crystallization, cure |
| Typical sample | 10–50 mg | 5–10 mg |
| Range at Covalent | Ambient to 1200 °C | −90 to 550 °C |
| Answers | “How much of it is there, and when does it break down?” | “When does it soften, melt or cure?” |
When to Run Both
Run both when a polymer behaves unexpectedly. DSC shows whether the Tg or crystallinity has shifted, and TGA shows whether the filler, plasticizer or moisture content has changed. Running TGA first also sets a safe upper temperature for the DSC run, so the sample doesn’t decompose inside the DSC cell.