Opens in a new tab

Seeing the Unseen: A Practical Guide to Electron Microscopy (SEM, TEM and STEM)

SEM, TEM and STEM each answer a different question about a material, and choosing the wrong one costs time, samples and clarity. Seeing the Unseen is a free 20-page eBook that explains how electron microscopy works, what each imaging mode reveals, how TEM samples are prepared by focused ion beam (FIB), and which settings change the image you get. It draws on real examples from semiconductor and electronics failure analysis, from PCB delamination to cracks in a CMOS gate..

SEM, TEM or STEM: which answers your question?

Each electron microscopy mode answers a different question. Picking the wrong one costs time, samples and clarity.

ModeWhat it showsSampleBest for
SEMSurface shape and texture, from mm down to ~1 nm; elements with EDSBulk parts, coated if non-conductiveFractures, particles, coatings, first look in failure analysis
TEMInternal structure down to individual atoms: crystal defects, interfaces, layer thicknessThin lamella, typically 20–100 nmThin films, grain boundaries, device cross-sections
STEMAtomic-number contrast plus element maps (EDS, EELS) at each pointSame lamella as TEMLayer chemistry, interfaces, dopant and composition mapping

How TEM samples are made

TEM and STEM need a slice thin enough for electrons to pass through, under 100 nm and often just tens of nanometers. A focused ion beam (FIB) cuts this lamella from the exact spot you care about, such as one transistor, via or defect:

  1. Cap the site. A platinum, carbon or tungsten layer protects the surface from the ion beam.
  2. Mill trenches. High-current ions cut away material on both sides, leaving a lamella about 1–2 µm thick.
  3. Undercut and lift out. A micromanipulator needle is welded on, and the lamella is cut free and mounted on a TEM grid.
  4. Thin and polish. The beam energy is stepped down, and a final 2–5 kV polish removes the damaged layer left by higher-energy milling.

Each TEM image represents hours of careful preparation, which is why prep quality decides both the result and the turnaround.

Settings that change your results

Many electron microscopy settings trade contrast against resolution. The eBook explains the ones that matter most:

  • Beam energy: low energy (around 1 kV) shows surface topography in SEM, while higher energy adds signal and reaches deeper. In TEM, higher energy improves resolution.
  • Probe size and beam current: a smaller probe gives sharper detail, while more current gives stronger contrast.
  • Sample thickness: thinner TEM lamellae reduce blur, until individual atomic columns become visible.
  • Dwell time: longer dwell lowers noise but slows imaging and can damage the sample.
  • Charging: why non-conductive samples smear or drift, and how coatings, grounding and lower beam settings fix it.

Examples in the eBook

  • PCB cross-section (BSE): delamination between copper plating and a pad, a cause of intermittent connections in high-reliability boards
  • Solder ball cross-section (BSE): composition contrast separating lead-tin solder, the copper microvia and epoxy
  • Unfilled vias (SE): checking the etch profile and cleanliness before metal fill
  • CMOS metal lines (STEM): grain structure and boundaries that affect resistance and voiding
  • CMOS gate (TEM): a crack likely caused by thermal or CMP stress
  • Bright field vs dark field STEM: the same device imaged both ways to reveal vias, liners, stacking faults and dislocations

Who it's for

Engineers and scientists in semiconductors, electronics, batteries, polymers and metals who order or interpret electron microscopy for process development, failure analysis or quality control.

Written by Valerie Brogden with Covalent’s electron microscopy team.